Appendix B: Waveform Data in MSO/DPO2000B and
MSO/DPO2000 Series Instruments
An analog channel of an MSO/DPO2000B and MSO/DPO2000 series
instrument has a sample waveform and a peak detect waveform being acquired
simultaneously. This combination waveform is referred to as a “composite”
waveform. In some cases, for example when averaging is turned on, or for math
waveforms
, a composite waveform is not available. In these cases, the waveform
is termed “singular”, and is available only in sample mode.
Full Versus Reduced
Resoluti
on Waveforms
The MSO/DPO2000/B series introduces the concept of reduced resolution
waveform
s available for curve queries. The MSO/DPO2000/B oscilloscopes
acquire channel data into “full” resolution records that are nominally 1 million
(1M) or 100,000 (100K) points in size (the acquisition record length); these are
termed “full resolution”, or LRL (long record length) waveforms. These LRL
records are “shrunk” (decimated) down to “ thumbnail”, “short”, or “reduced
resolution” waveforms; these are what appear on the oscilloscope’s display
scree
n. During decimation, a low-pass lter operation may be applied to the LRL
sample data that results in a ltered thumbnail wavefor m. In general, both full
and reduced resolution waveforms can be queried from the oscilloscope. The size
of the reduced resolution waveform varies from 800 to 6250 points, depending
on instrument settings.
Math FFT waveform data is available only in FULL resolution.
For applications which need to query waveform data rapidly for the purpose of
displaying it, reduced resolution waveform data will provide a faster display
update than a full resolution waveform data.
Nominal Versus Actual
Record Length
The MSO/DPO2000/B series also introduces the concept of nominal versus actual
acquisition record length. Like other oscilloscopes, the acquisition record length
can be set using the HORizontal:RECOrdlength command. However, the actual
number of points in the waveform record may differ from the nominal record
length established by the HORizontal:RECOrdlength setting, depending on the
HORizontal:SCAle in effect at the time of t he acquisition. The v alid values
for setting acquisition record length a re 1,000,000 (1M) and 100,000 (100K)
points. Depending on the h orizontal scale in effect at the time of the acquisition, a
nominal 1M point record may b e comprised of 1.0M or 1.25M (1,250,000) points
and a nominal 100K point record may be comprised of 100K or 125K (125,000)
points. The actual number of points in the waveform record for the source
waveform (specied by the DATa:SOUrce command) may be obtained using the
WFMOutpre:RECOrdlength? query.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual B-1