Commands Listed in Alphabetical Order
Arguments
<NR1> is a value
in the range from 1 to 10, which species a saved setup storage
location.
Examples
*RCL 3 restores the oscilloscope from a copy of the settings stored in memory
location 3.
RECAll:SETUp (No Query Form)
Restores the state of the oscilloscope from a copy of the s ettings stored in memory.
The settings are stored using the
*SAV command.
Group
Save and Recall
Syntax
RECAll:SETUp {FACtory|<NR1>|<file path>}
Related Commands
FACtory, *RCL, *RST, *SAV, SAVe:SETUp, FILESystem:CWD
Arguments
FACtory restores the factory setup.
<NR1> is a value in the range from 1 to 10, which species a saved setup storage
location.
<file p ath> species a location for an oscilloscope setup le. <le path> is
a quoted string that denes the le name and path. Input the le path using the
form
<drive>:/<dir>/<filename>. <extension> and one or <dir>s are
optional. If you do not specify them, the oscilloscope will read the le from the
default directory (see FILESystem:CWD). <lename> s tands for a lename; the
use of wildcard characters in lenames is not supported. Filename extensions are
not required, but highly recommended.
Examples
RECALL:SETUP FACTORY recalls (and makes current) the oscilloscope setup
to its factory defaults.
RECALL:SETUP 2 recalls the oscilloscope setup from setup storage location 2.
RECALL:SETUP "TEK00000.SET" recalls the setup from the le TEK00000.SET
in the default directory for setups (E:/tekScope/setups).
RECAll:WAVEform (No Query Form)
This command (no query form) recalls a stored waveform to a reference location.
2-228 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual