Connect-Tek USB-2 Spécifications Page 346

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 441
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 345
Commands Listed in Alphabetical Order
Examples
TRIGGER:A:PUL
SEWIDTH:WHEN LESST HAN
species that the duration of the A
pulse will fall within dened high and low limits.
TRIGGER:A:PU
LSEWIDTH:WHEN?
might return :TRIGGER:A:PULSEWIDTH:
WHEN TH AN
indicating that the conditions for generating a width trigger.
TRIGger:A:PULSEWidth:WIDth
Sets or returns the width setting for the pulse width trigger.
Group
Trigger
Syntax
TRIGger:A:PULSEWidth:WIDt h <NR3>
TRIGger:A:PULSEWidth:WIDt h?
Related Commands
TRIGger:A:PULSEWidth:WHEn
Arguments
<NR3> species the pulse width in seconds.
Examples
TRIGGER:A:PULSEWIDTH:WIDT H 5.0E-6 sets the pulse width to 5 µs.
TRIGGER:A:PULSEWIDTH:WIDT H? might return :TRIGGER:A:PULSEWIDTH:
WIDTH 2 .0000E-9
indicating that the pulse width is set to 2 ns.
TRIGger:A:RUNT? (Query Only)
Returns the current A runt trigger parameters.
Group
Trigger
Syntax
TRIGger:A:RUNT?
Examples
TRIGGER:A:RUNT? might return :TRIGGER:A:RUNT:SOURCE CH1;POLARIT Y
POSITIVE;WHEN OCCURS;WIDTH 4.0000E-9
.
TRIGger:A:RUNT:POLarity
Sets or returns the polarity for the runt trigger.