Connect-Tek USB-2 Spécifications Page 391

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 441
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 390
Commands Listed in Alphabetical Order
ZOOm:ZOOM1? (
Query Only)
Returns the current horizontal positioning and scaling of the display.
Group
Zoom
Syntax
ZOOm:ZOOM1?
Examples
ZOOM:ZOOM1? might return :ZOOM:ZOOM1:STAT E
1;SCALE 400.0000E-12;POSITION 46.8986;FACTOR
50.0000E+3;HORIZONTAL:POSI TION 46.8986;SCALE 400.0000E-12
.
ZOOm:ZOOM1:FACtor? (Query Only)
Returns the zoom factor of a particular zoom box.
Group
Zoom
Syntax
ZOOm:ZOOM1:FACtor?
Returns
<NR1> is the zoom factor of a z oom box.
ZOOm:ZOOM1:HO
Rizontal:POSition
Sets or returns the horizontal position for the specied zoom.
Group
Zoom
Syntax
ZOOm:ZOOM1:HORizontal:POSi tion <NR3>
ZOOm:ZOOM1:HORizontal:POSi tion?
Arguments
<NR3> is a value from 0 to 100.00 and is the percent of the upper window that is
to the left of screen center, when the zoom fac tor is or greater.
Examples
ZOOM:ZOOM1:HORIZONTAL:POSI TION 50 sets the Zoom1 reference pointer at
50% of acquired waveform.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-375
Vue de la page 390
1 2 ... 386 387 388 389 390 391 392 393 394 395 396 ... 440 441

Commentaires sur ces manuels

Pas de commentaire