Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:SETHold:CLOCk:TH Reshold {<NR3>|TTL|ECL}
TRIGger:A:SETHold:CLOCk:TH Reshold?
Arguments
TTL spec
ies a preset
TTL high level of 1.4 V.
ECL species a preset ECL high level of –1.3 V.
<NR3> is the clock level, in volts.
Examples
TRIGGER:A:SETHOLD:CLOCK:TH RESHOLD TTL species the preset TTL value
of 1.4 V as the clock threshold for the setup and hold trigger.
TRIGGER:A:SETHOLD:CLOCK:TH RESHOLD? might return
:TRIGGER:A:LOGIC:SETHOLD:C LOCK:THRESHOLD 1.2000E+00
indicating that the clock threshold for the setup and hold trigger is 1.2 V.
TRIGger:A:SETHold:DATa? (Query Only)
Returns the voltage threshold and data source for the setup and hold trigger.
Group
Trigger
Syntax
TRIGger:A:SETHold:DATa?
Related Commands
TRIGger:A:SETHold:CLOCk?
Examples
TRIGGER:A:SETHOLD:DATA? might return
:TRIGGER:A:SETHOLD:DATA:SO URCE CH2;THRESHOLD 80.0000E-3
TRIGger:A:SETHold:DATa:SOUrce
Sets or returns the data source for the setup and hold trigger. You cannot specify
the same source for both clock and d
ata.
For DPO models, you can specify only a single data source. Data sources for
DPO models may be one of CH1–CH4 or
D0–D15.
Group
Trigger
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-335