Commands Listed in Alphabetical Order
Examples
SELECT:CH2 ON t
urns the channel 2 waveform display on, and selects channel 2.
SELECT:CH1? might return :SELECT:CH1 1 indicating that channel 1 is being
displayed.
SELect:CONTROl
Sets or returns the waveform that is the recipient of future channel-related
commands, for example, the cursor commands. The command form also performs
the equivalent of a SELect:CH<x> ON command, as well as the Math, Reference,
and Bus variations of that command.
Group
Vertical
Syntax
SELect:CONTROl {CH<x>|MATH|BUS<x>}
SELect:CONTROl?
Arguments
CH<x> species a channel waveform as the waveform affe
cted by the front-panel
controls. <x> is the channel number.
MATH species the math waveform as the waveform that i
saffectedbythe
front-panel controls.
BUS<x> species a bus waveform as the waveform affec
tedbythefront-panel
controls. <x> species the bus number.
Returns
NONE if all the channels are turned off. NONE is ignored on input.
Examples
SELECT:CONTROL CH2 resets acquisition displays on channel 2, and causes the
selected waveform to be the implied object of waveform commands.
SELECT:CONTROL? might return :SELECT:CONTROL MATH indicating that math
is the implied object of waveform commands.
SELect:D<x>
Turns on the display of the digital channel <x> and resets the acquisition. <x > is
the channel number. The query returns whether the channel is on or o
ff but does
not indicate whether it is the selected waveform.
Group
Vertical
2-286 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual