Commands Listed in Alphabetical Order
Arguments
ENV spec ies th
at the waveform is transmitted in envelope mode as maximum
and minimum point pairs. Only
Y values are explicitly transmitted. Absolute
coordinates are given by:
Xn = XZEro + XINcr (n PT_Off)
Ynmax = YZEr
o+YMUlt(ynmax–YOFf)
Ynmin = YZEro + YMUlt (ynmin YOFf)
Y speci es a normal waveform where one ASCII or binary data point is
transmitted for each point in the waveform record. Only
Y values are explicitly
transmitt
ed. Absolute coordinates are given by:
Xn = XZEro + XINcr (n PT_Off)
Yn = YZEro + YMUlt (yn YOFf)
Examples
WFM
INPRE:PT_FMT ENV
sets the incoming w aveform data point format to
enveloped.
WF
MINPRE:PT_FMT?
might return :WFMINPRE:PT_FMT ENV indicating that the
waveform is transmitted as maximum and minimum point pairs.
WFMInpre:PT_Off
The set form of this command is ignored. The q uery form always returns a 0. This
command is listed for compatibility with other Tektronix oscilloscopes.
Group
Waveform Transfer
Syntax
WFMInpre:PT_Off <NR1>
WFMInpre:PT_Off?
Arguments
Arguments are ignored.
WFMInpre:XINcr
Sets or returns the horizontal interval between incoming waveform points in
units specied by WFMInpre:XUNit.
Group
Waveform Transfer
2-358 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual