Commands Listed in Alphabetical Order
Arguments
ECL species a p
reset
ECL high level of –1.3 V.
TTL species a TTL preset high level of 1.4 V.
<NR3> species the threshold level in volts.
BUS:THReshold:D<x>
Sets or retu
rns the threshold for digital channel <x>, where x is the digital channel
number. This will apply to all search and trigger types that use the channel. The
threshold value of D<x> is set t o the closest value used by hardware.
Digital channel thresholds are set in two groups. One for channels D0–D7 and the
other for D8–D15. Setting the thresholds in one of these groups sets the threshold
for all the channels in that group.
Conditions
This command requires a DPO2AUTO, DPO2EMBD, or D PO2COMP application
module.
Group
Bus
Syntax
BUS:THReshold:D<x> {<NR3>|ECL|TTL}
BUS:THReshold:D<x>?
Related Commands
TRIGger:A:LEVel:D<x>
Arguments
ECL species a preset ECL high level of –1.3 V.
TTL species a preset TTL high level of 1.4 V.
<NR3> species the threshold level in volts.
BUS:UPPerthreshold:CH<x>
Sets the upper threshold for each channel. This applies to a ll search and trigger
types that use the channel.
Group
Bus
Syntax
BUS:UPPerthreshold:CH<x> {<NR3>|ECL|TT L}
BUS:UPPerthreshold:CH<x>?
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-93