Commands Listed in Alphabetical Order
CALibrate:RE
Sults:FACtory? (Query Only)
Returns the status of internal and factory calibration, without performing any
calibration operations.
Group
Calibration a nd Diagnostic
Syntax
CALibrate:RESults:FACtory ?
CALibrate:RESults:SPC? (Query Only)
Returns the status of the SPC operation. This query does not initiate a SPC.
Group
Calibration a nd Diagnostic
Syntax
CALibrate:RESults:SPC?
Related Commands
*CAL?
Returns
INIT indicates that SPC has never suc cessfully completed.
PASS indicates that the last SPC operation passed.
FAIL indicates that the last SPC operation failed.
RUNNING indicates that the SPC operation is running.
Examples
CALIBRATE:RESULTS:SPC? returns the results of the last SPC operation: either
PASS or FAIL.
CALibrate:TEMPerature? (Query Only)
Returns 0 as the MSO/DPO2000/B series instruments do not support recording
of oscilloscope temperature.
Group
Calibration a nd Diagnostic
Syntax
CALibrate:TEMPerature?
2-98 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series O scilloscopes Programmer Manual