Connect-Tek USB-2 Spécifications Page 23

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 441
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 22
Command Syntax
Measurement Specier
Mnemonics
Commands can sp
ecify whic h measurement to set or query as a mnemonic in the
header. Up to four automated measurements may be displayed.
Table 2-8: Measurement Specier Mnemonics
Symbol Meaning
MEAS<x> A measurement specier; <x> is 1 through 4.
Channel Mnemonics
Commands specify the channel to use as a mnemonic in the header.
Reference Waveform
Mnemonics
Commands can specify the reference waveform to use as a mnemonic in the
header.
Table 2-9: Reference Waveform Mnemonics
Symbol Meaning
REF<x>
A reference waveform specier; <x> 1 or 2. The MSO/DPO2000B and
MSO/DPO2000 series provides only two REF waveforms regardless of
whether the instrument is a 2 or 4 channel model.
Argument Types
Numeric
Ma
ny oscilloscope commands require numeric arguments. The syntax shows
the format that the oscilloscope returns in response to a query. This is also the
preferred format when sending the command to the o scilloscope though any of
the formats will be accepted. This documentation represents these arguments as
described below.
Table 2-10: Numeric Arguments
Symbol Meaning
<NR1>
Signed integer value
<NR2> Floating point value without an exponent
<NR3> Floating point value with a n exponent
<bin>
Digital data in binary format
Most num
eric arguments will be automatically forced to a valid setting, by either
rounding or truncating,, when an invalid number is input, unless otherwise noted
in the command description.
Quoted String
Some commands accept or return data in the form of a quoted string, which is
simply a group of ASCII characters enclosed by a single quote (') or double quote
("). The following is an example of a quoted string:
"This i s a quote d
string"
. This documentation represents these arguments as follows:
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-7