Connect-Tek USB-2 Spécifications Page 349

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 441
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 348
Commands Listed in Alphabetical Order
Examples
TRIGGER:A:RUN
T:WIDTH 15E-6
sets the minimum width of the pulse runt
triggerto1s.
TRIGGER:A:RU
NT:WIDTH?
might return :TRIGGER:A: PULSE:RUNT:WIDTH
2.0000E-09
indicating that the minimum width of a pulse runt trigger is 2 ns.
TRIGger:A:SETHold? (Query Only)
Returns the clock edge polarity, voltage threshold and source input; data voltage
threshold and source; and both setup and hold times for setup and hold violation
triggering.
Group
Trigger
Syntax
TRIGger:A:SETHold?
Examples
TRIGGER:A:SETHOLD? might return
:TRIGGER:A:SETHOLD:CLOCK:S OURCE CH1;EDGE RISE;THRESHOLD
100.0000E-3;:TRIGGER:A:SET HOLD:DATA:SOURCE CH2;THRE SHOLD
80.0000E-3;:TRIGGER:A:SETH OLD:HOLDTIME 20.0000E-9;SETT IME
8.0000E-9
TRIGger:A:SETHold:CLOCk? (Query Only)
Returns the clock edge polarity, voltage threshold, and source input for setup
and hold triggering.
Group
Trigger
Syntax
TRIGger:A:SETHold:CLOCk
?
Examples
TRIGGER:A:SETHOLD:CLOCK? might return
:TRIGGER:A:SETHOLD:CLOCK:S OURCE EXT;EDGE FALL;THRESHOLD
1.4000
TRIGger:A:SETHold:CLOCk:EDGE
Sets or returns the clock edge polarity for setup and hold triggering.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-333