Connect-Tek USB-2 Spécifications Page 112

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 441
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 111
Commands Listed in Alphabetical Order
CALibrate:FA
Ctory:STATus? (Query Only)
Returns the factory calibration status value saved in nonvolatile memory.
Group
Calibration a nd Diagnostic
Syntax
CALibrate:FACtory:STATus?
Examples
CALIBRATE:FACTORY:STATUS? might return CALI BRATE:FACTORY:ST ATUS
PASS
indicating that factory calibration passed.
CALibrate:INTERNal (No Query Form)
This command starts a signal path compensation.
Group
Calibration a nd Diagnostic
Syntax
CALibrate:INTERNal
Arguments
None
Examples
CALIBRATE:INTERNAL starts a serial path compensation c ycle.
CALibrate:INTERNal:STARt (No Query Form)
Starts the internal signal path calibration (SPC) of the oscilloscope. You can
use the CALibrate:INTERNal:STATus? query to return the current status of the
internal signal path calibration of the oscilloscope.
Group
Calibration a nd Diagnostic
Syntax
CALibrate:INTERNal:STARt
Related Commands
CALibrate:RESults:SPC?
Examples
CALIBRATE:INTERNAL:START initiates the internal signal path calibration of
the oscilloscope.
2-96 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual
Vue de la page 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 440 441

Commentaires sur ces manuels

Pas de commentaire