Calibration can be accomplished either through the use of a resistance
standard that represents the load or the actual load itself. The range of
acceptable loop resistance is in the Specifications section. A calibrated,
high resolution Voltmeter should be connected across the resistance or a
high-resolution milli-ammeter connected in series with the load. When
using a voltmeter, calculate the current with the formula I = E/R.
Click on the Analog Output Calibration tab. Click Start to initiate the calibration
sequence.
e box is highlighted, type in
x. Once the uncalibrated values
ar me active.
Each of the boxes in the Uncalibrated Values section will be highlighted in
sequence, starting with Positive Maximum. When th
the measured value in mA and Tab to the next bo
e taken, the Calibrated Values section will beco
71